Class II Recall

GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diff

Greer Laboratories, Inc.

Summary

The FDA issued a Class II for GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSI by Greer Laboratories, Inc.. Reason: The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative res.

Details

Source

Device Recall

External ID

Z-0347-2023

Action Date

2022-12-07

Status

Terminated

Category

device

Product Description

GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.

Lot/Code Info: Lot Number 389591, exp 6/12/2023

Quantity Affected: 218 units

Reason for Recall

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.

Distribution

US Distribution to states of: CA, CO, IA, KY, MA, MD, MO, NC, PA, UT, VA, and International to countries of: Canada, Denmark

Type: Voluntary: Firm initiated

Recall Initiated: 2022-10-25

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 164 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Greer Laboratories, Inc. has 2 FDA actions in our database, including 2 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Greer Laboratories, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Greer Laboratories, Inc. have FDA actions?

Greer Laboratories, Inc. has 2 FDA actions in our database, including 2 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0347-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions