Summary
The FDA issued a Class II for Mobile column TruSystem 7500 U, Product Code 1730720 by Baxter Healthcare Corporation. Reason: The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the.
Details
Source
Device Recall
External ID
Z-0333-2027
Action Date
2026-11-11
Status
Ongoing
Category
device
Product Description
Mobile column TruSystem 7500 U, Product Code 1730720
Lot/Code Info: UDI-DI: 00887761974388; Serial number: 110183858
Quantity Affected: 1 unit
Reason for Recall
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Distribution
US Nationwide distribution in the states of IL, MO, TX.
Type: Voluntary: Firm initiated
Recall Initiated: 2026-09-14
Company
Round Lake, IL
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 127 device recalls issued in the same week, part of 315 device-related FDA actions this month.
Baxter Healthcare Corporation has 340 FDA actions in our database, including 318 recalls and 22 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Baxter Healthcare Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Baxter Healthcare Corporation have FDA actions?
Baxter Healthcare Corporation has 340 FDA actions in our database, including 318 recalls and 22 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0333-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29