RecallHawk
Class II Recall

Precision Pointer Electromagnetic (8000-050-001)

Stryker Leibinger GmbH & Co. KG

Summary

The FDA issued a Class II for Precision Pointer Electromagnetic (8000-050-001) by Stryker Leibinger GmbH & Co. KG. Reason: Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the devic.

Details

Source

Device Recall

External ID

Z-0314-2027

Action Date

2026-11-04

Status

Ongoing

Category

device

Product Description

Precision Pointer Electromagnetic (8000-050-001)

Lot/Code Info: Lot 7000413361 UDI-DI 07613327413205

Quantity Affected: 513

Reason for Recall

Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.

Distribution

Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-24

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 202 device recalls issued in the same week, part of 315 device-related FDA actions this month.

Stryker Leibinger GmbH & Co. KG has 19 FDA actions in our database, including 5 recalls and 14 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Stryker Leibinger GmbH & Co. KG) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Stryker Leibinger GmbH & Co. KG have FDA actions?

Stryker Leibinger GmbH & Co. KG has 19 FDA actions in our database, including 5 recalls and 14 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0314-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.