Summary
The FDA issued a Class II for DeRoyal CHILDREN'S GENERAL TRACEPACK, REF 89-9342-06 by DeRoyal Industries Inc. Reason: DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Str.
Details
Source
Device Recall
External ID
Z-0307-2025
Action Date
2024-11-13
Status
Ongoing
Category
device
Product Description
DeRoyal CHILDREN'S GENERAL TRACEPACK, REF 89-9342-06
Lot/Code Info: UDI/ DI 00749756916646 (each), 50749756916641 (case): Lot Numbers: lot 59839716 exp 12/1/2025
Quantity Affected: 160 units
Reason for Recall
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
Distribution
US Nationwide distribution in the state of OH.
Type: Voluntary: Firm initiated
Recall Initiated: 2024-10-03
Company
Powell, TN
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 145 device recalls issued in the same week, part of 403 device-related FDA actions this month.
DeRoyal Industries Inc has 183 FDA actions in our database, including 181 recalls and 2 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (DeRoyal Industries Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does DeRoyal Industries Inc have FDA actions?
DeRoyal Industries Inc has 183 FDA actions in our database, including 181 recalls and 2 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0307-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29