RecallHawk
Class II Recall

Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL BLADE, REF 73-8020, STERILE EO

Southmedic, Inc.

Summary

The FDA issued a Class II for Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL BLADE, REF 73-8020, ST by Southmedic, Inc.. Reason: Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to d.

Details

Source

Device Recall

External ID

Z-0279-2023

Action Date

2022-11-30

Status

Ongoing

Category

device

Product Description

Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL BLADE, REF 73-8020, STERILE EO

Lot/Code Info: GTIN 620974002628, Lot number: W92761

Quantity Affected: 450 units

Reason for Recall

Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to distribution where it was sold and distributed to customers as non-sterile product labelled as sterile.

Distribution

US Nationwide distribution in the state of IL.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-10-27

Company

Southmedic, Inc.

Barrie, N/A

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 152 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Southmedic, Inc. has 3 FDA actions in our database, including 2 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Southmedic, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Southmedic, Inc. have FDA actions?

Southmedic, Inc. has 3 FDA actions in our database, including 2 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0279-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.