RecallHawk
Class II Recall

Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - TruFlight Select PET/CT System Number: 882438 Con

Philips North America Llc

Summary

The FDA issued a Class II for Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - TruFlight S by Philips North America Llc. Reason: Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Ar.

Details

Source

Device Recall

External ID

Z-0279-2022

Action Date

2021-12-01

Status

Ongoing

Category

device

Product Description

Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - TruFlight Select PET/CT System Number: 882438 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.

Lot/Code Info: Serial Number: 17006

Quantity Affected: 1 unit

Reason for Recall

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part

Distribution

US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to countries of: Australia, Germany, Netherlands and Saudi Arabia.

Type: Voluntary: Firm initiated

Recall Initiated: 2021-05-01

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 132 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Philips North America Llc has 302 FDA actions in our database, including 302 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Philips North America Llc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Philips North America Llc have FDA actions?

Philips North America Llc has 302 FDA actions in our database, including 302 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0279-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.