Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

Abbott Molecular, Inc.

Summary

The FDA issued a Class II for Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use by Abbott Molecular, Inc.. Reason: Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flappe.

Details

Source

Device Recall

External ID

Z-0268-2023

Action Date

2022-11-30

Status

Ongoing

Category

device

Product Description

Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

Lot/Code Info: All serial numbers

Quantity Affected: 883 units

Reason for Recall

Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and locked. It was found that the software will cause the flapper door to stay in the closed position when the drawer is closed and locked if the flapper had previously been moved out of position. 2) Sample preparation drawer #1 barcode information is used instead of sample prep drawer #2 when the scanned data is not sent to the System Control Center before the next bottle barcode in sample prep drawer#2 is scanned. 3) Under a specific condition, while the system is processing tests and a new test request is made, when the level of the bulk solution is too low to process a test, the software would error stop the system and try to complete all in-process tests. An internal counter is reset which can cause reaction vessels (RVs) already present in an Amplification Detection Unit (or amp detect) to remain there and not be moved to waste so that more RVs can be placed into the amp detect for processing. 4) 4 to 6 replicates of the same auto calibrator orders can be run with 2 different sets of calibrator materials on the same rack. It was discovered during internal testing, when running 4 to 6 replicates of a calibrator, it is possible for the user to use 2 separate lots of material which is not detected by the System Control Center (SCC). If this were to occur, the calibration curve would be created using 2 separate material lots. In normal use scenario, this curve would be made using only 1 lot of material.

Distribution

Worldwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-10-12

Company

Abbott Molecular, Inc.

Des Plaines, IL

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 152 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Abbott Molecular, Inc. has 39 FDA actions in our database, including 30 recalls and 9 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Abbott Molecular, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Abbott Molecular, Inc. have FDA actions?

Abbott Molecular, Inc. has 39 FDA actions in our database, including 30 recalls and 9 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0268-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions