RecallHawk
Class II Recall

Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular syst

Spectranetics Corporation

Summary

The FDA issued a Class II for Philips Laser System, REF: LAS-100. Used in minimally invasive interventional p by Spectranetics Corporation. Reason: Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be ab.

Details

Source

Device Recall

External ID

Z-0221-2023

Action Date

2022-11-23

Status

Ongoing

Category

device

Product Description

Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular system, and for the removal of problematic pacemaker and defibrillator cardiac leads.

Lot/Code Info: UDI-00813132027452, Software Version: 0.5.0.3, Serial Numbers: 100000, 100001, 100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100012, 100013, 100014, 100015, 100016, 100017, 100018, 100019, 100020, 100021, 100022, 100023, 100024, 100025, 100026, 100027, 100028, 100029, 100030, 100033, 100034, 100035, 100036, 100037, 100038, 100039, 100040, 100041, 100042, 100043, 100044, 100045, 100046, 100047, 100048, 100049, 100050, 100051, 100052, 100053, 100054, 100055, 100058, 100059, 100060, 100061, 100062, 100063, 100064, 100065, 100066, 100067, 100068, 100069, 100070, 100071, 100072, 100073, 100074, 100075, 100076, 100077, 100078, 100079, 100080, 100081, 100082, 100083, 100084, 100085, 100086, 100087, 100088, 100089, 100090, 100091, 100092, 100093, 100094, 100095, 100096, 100097, 100098, 100099, 100100, 100101, 100102, 100103, 100104, 100105, 100106, 100107, 100108, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100120

Quantity Affected: 113

Reason for Recall

Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation of treatment delay and/or treatment beyond initial scope and/or unable to treat patient.

Distribution

US: LA, MO, TN, CA, NC, PA, HI, AZ, CO, OK, FL, TX, MN, MI, VA, CT, NY, MD, IL, NJ, MA, KS, ME, AR, IA, IN, GA, ID, OR. OUS: DE, GB, ES, SE, IT, RS, GR, IN, FR, IL, DK, SA

Type: Voluntary: Firm initiated

Recall Initiated: 2022-10-14

Company

Spectranetics Corporation

Colorado Springs, CO

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 171 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Spectranetics Corporation has 5 FDA actions in our database, including 5 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Spectranetics Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Spectranetics Corporation have FDA actions?

Spectranetics Corporation has 5 FDA actions in our database, including 5 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0221-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions