RecallHawk
Class II Recall

ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Mo

Beckman Coulter, Inc.

Summary

The FDA issued a Class II for ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the f by Beckman Coulter, Inc.. Reason: Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that .

Details

Source

Device Recall

External ID

Z-0202-2027

Action Date

2026-10-28

Status

Ongoing

Category

device

Product Description

ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)

Lot/Code Info: Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860

Quantity Affected: 336 units

Reason for Recall

Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-21

Company

Beckman Coulter, Inc.

Indianapolis, IN

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 187 device recalls issued in the same week, part of 480 device-related FDA actions this month.

Beckman Coulter, Inc. has 155 FDA actions in our database, including 111 recalls and 44 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Beckman Coulter, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Beckman Coulter, Inc. have FDA actions?

Beckman Coulter, Inc. has 155 FDA actions in our database, including 111 recalls and 44 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0202-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.