RecallHawk
Class II Recall

Insulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711, MMT-1712, MMT-1751, MMT-1752

Medtronic MiniMed

Summary

The FDA issued a Class II for Insulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711, MMT-1712, M by Medtronic MiniMed. Reason: Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that coul.

Details

Source

Device Recall

External ID

Z-0194-2023

Action Date

2022-11-16

Status

Ongoing

Category

device

Product Description

Insulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711, MMT-1712, MMT-1751, MMT-1752

Lot/Code Info: Model/UDI-GTIN (All Serial Numbers): MMT-1750/00763000375461, 00643169559745; MMT-1711/643169554931, 763000367039, 00643169890039, 00643169554917, 00643169629813, 00643169554924, 00643169554931, 00643169742062, 00763000013066, 00643169554948, 00643169554955, 00643169621954; MMT-1712/643169662612, 643169577701, 00763000205409, 00643169890046, 00643169577664, 00643169629820, 00643169577671, 00643169577688, 00763000155346, 00643169577695, 00643169577701, 00643169621961; MMT-1751/643169672239, 00643169574410, 00643169574427, 00643169574434, 00763000253288, 00643169574441, 00643169574458, 00643169643512, 00643169521155, 00643169521704, 00763000015596, 00643169520882, 00643169521421, 00643169521971; MMT-1752/763000192181, 00643169596368, 00643169596382, 00643169596405, 00763000318291, 00763000318307, 00643169596443, 00643169522305, 00643169521292, 00643169522381, 00643169521841, 00643169522268, 00643169521025, 00643169522343, 00643169521575, 00643169522114

Quantity Affected: 316326

Reason for Recall

Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.

Distribution

US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: Andorra, Australia, Austria, Barbados, Belgium, Bermuda, Bolivia, Botswana, Canada, Canary Islands, Cayman Islands, Chile, Czech Republic, Denmark, Dominican Rep., El Salvador, Faroe Islands, Finland, France, Germany, Gibraltar, Greece, Guatemala, Hungary, Ireland, Israel, Italy, Kenya, Liechtenstein, Luxembourg, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Paraguay, Peru, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Trinidad and Tobago, Turkey, Uganda, United Kingdom, Venezuela

Type: Voluntary: Firm initiated

Recall Initiated: 2022-09-20

Company

Medtronic MiniMed

Northridge, CA

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 134 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Medtronic MiniMed has 16 FDA actions in our database, including 12 recalls and 4 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medtronic MiniMed) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medtronic MiniMed have FDA actions?

Medtronic MiniMed has 16 FDA actions in our database, including 12 recalls and 4 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0194-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions