RecallHawk
Class II Recall

Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1

Siemens Healthcare Diagnostics, Inc.

Summary

The FDA issued a Class II for Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Cent by Siemens Healthcare Diagnostics, Inc.. Reason: The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sa.

Details

Source

Device Recall

External ID

Z-0193-2022

Action Date

2021-11-10

Status

Ongoing

Category

device

Product Description

Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1 and lower. SMN 10703038

Lot/Code Info: UDI 00630414596754

Quantity Affected: 1142

Reason for Recall

The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical interpretations and clinical decisions.

Distribution

Worldwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2021-10-06

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 120 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Siemens Healthcare Diagnostics, Inc. has 166 FDA actions in our database, including 142 recalls and 24 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Siemens Healthcare Diagnostics, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Siemens Healthcare Diagnostics, Inc. have FDA actions?

Siemens Healthcare Diagnostics, Inc. has 166 FDA actions in our database, including 142 recalls and 24 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0193-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions