Medline surgical convenience kits labeled as: 1) EAR, Medline product number/SKU DYNJ909020B; 2) EAR, Medline produ
Summary
The FDA issued a Class II for Medline surgical convenience kits labeled as: 1) EAR, Medline product number/ by Medline Industries, LP. Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribut.
Details
Source
Device Recall
External ID
Z-0173-2027
Action Date
2026-10-21
Status
Ongoing
Category
device
Product Description
Medline surgical convenience kits labeled as: 1) EAR, Medline product number/SKU DYNJ909020B; 2) EAR, Medline product number/SKU DYNJ909020A; 3) CAROTID PACK, Medline product number/SKU DYNJ42922C; 4) EAR, Medline product number/SKU DYNJ909020; 5) FEM CAROTID PACK, Medline product number/SKU DYNJ65706; 6) CAROTID CDS-LF, Medline product number/SKU CDS983139B.
Lot/Code Info: Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 26AMF311; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25JMD500; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25DMJ849; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25CMD823; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25BMG250; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 24LMG633; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24KMB736; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24JMD634; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24IMD605; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 24HBT243; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 24HBD143; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24GMG103; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24DMI379; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 24CBJ928; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 24BME929; Medline product number/SKU DYNJ65706, UDI/DI 10193489284904 (each) 40193489284905 (case), Lot number 24AMF210; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 24AMD338; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 23LME666; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 23LMB230; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 23KMF895; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 23KMD188; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 23HMH838; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 23HBG421; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 23FBD203; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 23EBL260; Medline product number/SKU DYNJ909020, UDI/DI 10195327248673 (each) 40195327248674 (case), Lot number 23CMB011; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 23CBU833; Medline product number/SKU CDS983139B, UDI/DI 10193489211313 (each) 40193489211314 (case), Lot number 22BBV592; Medline product number/SKU CDS983139B, UDI/DI 10193489211313 (each) 40193489211314 (case), Lot number 22BBJ850; Medline product number/SKU CDS983139B, UDI/DI 10193489211313 (each) 40193489211314 (case), Lot number 22ABL217; Medline product number/SKU CDS983139B, UDI/DI 10193489211313 (each) 40193489211314 (case), Lot number 21LBM269; Medline product number/SKU CDS983139B, UDI/DI 10193489211313 (each) 40193489211314 (case), Lot number 21HBC597; Medline product number/SKU CDS983139B, UDI/DI 10193489211313 (each) 40193489211314 (case), Lot number 21GBE864; Medline product number/SKU CDS983139B, UDI/DI 10193489211313 (each) 40193489211314 (case), Lot number 21FBJ082.
Quantity Affected: 620 kits
Reason for Recall
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Distribution
US Nationwide distribution.
Type: Voluntary: Firm initiated
Recall Initiated: 2026-09-02
Company
Northfield, IL
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 150 device recalls issued in the same week, part of 446 device-related FDA actions this month.
Medline Industries, LP has 607 FDA actions in our database, including 576 recalls and 31 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Medline Industries, LP have FDA actions?
Medline Industries, LP has 607 FDA actions in our database, including 576 recalls and 31 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-0173-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29