RecallHawk
Class II Recall

THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrop

Johnson & Johnson Surgical Vision, Inc.

Summary

The FDA issued a Class II for THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. by Johnson & Johnson Surgical Vision, Inc.. Reason: Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) .

Details

Source

Device Recall

External ID

Z-0169-2023

Action Date

2022-11-09

Status

Ongoing

Category

device

Product Description

THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.

Lot/Code Info: UDI-DI/Lots: 10846835010183/30779056L, 30808001L, 30814782L, 30814783L, 30779172L

Quantity Affected: 91

Reason for Recall

Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.

Distribution

Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-09-07

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 124 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Johnson & Johnson Surgical Vision, Inc. has 13 FDA actions in our database, including 12 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Johnson & Johnson Surgical Vision, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Johnson & Johnson Surgical Vision, Inc. have FDA actions?

Johnson & Johnson Surgical Vision, Inc. has 13 FDA actions in our database, including 12 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0169-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions