RecallHawk
Class II Recall

Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.

Howmedica Osteonics Corp.

Summary

The FDA issued a Class II for Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514. by Howmedica Osteonics Corp.. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is .

Details

Source

Device Recall

External ID

Z-0166-2027

Action Date

2026-10-21

Status

Ongoing

Category

device

Product Description

Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.

Lot/Code Info: Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

Quantity Affected: 359 trays

Reason for Recall

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Distribution

These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-16

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 150 device recalls issued in the same week, part of 446 device-related FDA actions this month.

Howmedica Osteonics Corp. has 128 FDA actions in our database, including 128 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Howmedica Osteonics Corp.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Howmedica Osteonics Corp. have FDA actions?

Howmedica Osteonics Corp. has 128 FDA actions in our database, including 128 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0166-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions