RecallHawk
Class II Recall

PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H

Ethicon, LLC

Summary

The FDA issued a Class II for PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, by Ethicon, LLC. Reason: Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product desi.

Details

Source

Device Recall

External ID

Z-0153-2027

Action Date

2026-10-21

Status

Ongoing

Category

device

Product Description

PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H

Lot/Code Info: UDI-DI:Box/Unit: 30705031047369/10705031047365. Lot: RLBDPJ. Expiration: SEPTEMBER 30, 2026

Quantity Affected: 1764

Reason for Recall

Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.

Distribution

Worldwide - US Nationwide distribution in the states of SC, PA, TX, CA, ID, LA, MO, AL, NC, UT, IL, MA, OH, AZ, TN, OK, NE, ME, IN, WI, FL, CO, WY, NH, NY, NC, MN and the country of CA.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-08

Company

Ethicon, LLC

San Lorenzo, PR

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 150 device recalls issued in the same week, part of 446 device-related FDA actions this month.

Ethicon, LLC has 7 FDA actions in our database, including 7 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Ethicon, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Ethicon, LLC have FDA actions?

Ethicon, LLC has 7 FDA actions in our database, including 7 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0153-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.