Class I Recall

VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.

Ventec Life Systems, Inc.

Summary

The FDA issued a Class I for VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for by Ventec Life Systems, Inc.. Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of v.

Details

Source

Device Recall

External ID

Z-0127-2027

Action Date

2026-10-28

Status

Ongoing

Category

device

Product Description

VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.

Lot/Code Info: All Serial Numbers: UDI: 00850018761130

Quantity Affected: 100 units

Reason for Recall

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Distribution

Distribution US nationwide including Puerto Rico.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-09-17

Company

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 187 device recalls issued in the same week, part of 480 device-related FDA actions this month.

Ventec Life Systems, Inc. has 19 FDA actions in our database, including 19 recalls.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Ventec Life Systems, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Ventec Life Systems, Inc. have FDA actions?

Ventec Life Systems, Inc. has 19 FDA actions in our database, including 19 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0127-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.