RecallHawk
Class II Recall

Alphenix INFX-8000C/B5

Canon Medical System, USA, INC.

Summary

The FDA issued a Class II for Alphenix INFX-8000C/B5 by Canon Medical System, USA, INC.. Reason: Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalig.

Details

Source

Device Recall

External ID

Z-0108-2027

Action Date

2026-10-14

Status

Ongoing

Category

device

Product Description

Alphenix INFX-8000C/B5

Lot/Code Info: Serial no. B5F2532011/UDI: 4987670100192

Quantity Affected: 1 unit

Reason for Recall

Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error

Distribution

US Nationwide distribution in the state of ID.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-08-26

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 220 device recalls issued in the same week, part of 502 device-related FDA actions this month.

Canon Medical System, USA, INC. has 23 FDA actions in our database, including 23 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Canon Medical System, USA, INC.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Canon Medical System, USA, INC. have FDA actions?

Canon Medical System, USA, INC. has 23 FDA actions in our database, including 23 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0108-2027" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.