RecallHawk
Class II Recall

Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended

B. Braun Medical, Inc.

Summary

The FDA issued a Class II for Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carto by B. Braun Medical, Inc.. Reason: The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The is.

Details

Source

Device Recall

External ID

Z-0091-2023

Action Date

2022-10-26

Status

Ongoing

Category

device

Product Description

Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector.

Lot/Code Info: UDI-DI (GTIN): 04046964540943 Lots/Expiration Dates 61705083 31-Oct-22 61705124 31-Oct-22 61706055 31-Oct-22 61714103 31-Dec-22 61719696 31-Jan-23 61732456 31-Mar-23 61736580 31-May-23 61739427 31-May-23 61744671 31-Jul-23 61747367 31-Aug-23 61750206 31-Aug-23 61752658 30-Sep-23 61761422 30-Nov-23 61776447 30-Apr-24 61780880 30-Apr-24 61802320 30-Sep-24 61806067 31-Oct-24 61807459 31-Oct-24 61822282 28-Feb-25

Quantity Affected: 354,200 US; 1,800 OUS

Reason for Recall

The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

Distribution

Distribution US nationwide, Canada and Saudi Arabia.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-09-15

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 184 device recalls issued in the same week, part of 403 device-related FDA actions this month.

B. Braun Medical, Inc. has 282 FDA actions in our database, including 269 recalls and 13 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (B. Braun Medical, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does B. Braun Medical, Inc. have FDA actions?

B. Braun Medical, Inc. has 282 FDA actions in our database, including 269 recalls and 13 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0091-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions