RecallHawk
Class II Recall

CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with hea

CardioQuip, LLC

Summary

The FDA issued a Class II for CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RM by CardioQuip, LLC. Reason: There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result.

Details

Source

Device Recall

External ID

Z-0069-2022

Action Date

2021-10-13

Status

Ongoing

Category

device

Product Description

CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.

Lot/Code Info: All Serial Numbers

Quantity Affected: 1380 units

Reason for Recall

There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery which could result in patient infection.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of TX, PA, FL, CA, HI, SC, OH, AK, NY, IL, AR, WI, MI, AL, KS, NJ, GA, AZ, KY, MS, TN, LA, DE, MA, VA, CO, NE, NC, NH, ME, WV, PA, MO, MN, IN, CT, NH, NV, UT and the country of Venezuela.

Type: Voluntary: Firm initiated

Recall Initiated: 2021-06-24

Company

CardioQuip, LLC

College Station, TX

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 192 device recalls issued in the same week, part of 403 device-related FDA actions this month.

CardioQuip, LLC has 9 FDA actions in our database, including 9 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (CardioQuip, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does CardioQuip, LLC have FDA actions?

CardioQuip, LLC has 9 FDA actions in our database, including 9 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0069-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions