Class I Recall

KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. Jaco re

Kinova Inc

Summary

The FDA issued a Class I for KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 by Kinova Inc. Reason: Potential installation issue that could result in a fire hazard in the event of an interaction between a damaged Jaco arm and a wheelchair exhibiting .

Details

Source

Device Recall

External ID

Z-0049-2025

Action Date

2024-10-23

Status

Ongoing

Category

device

Product Description

KINOVA Jaco assistive robot. Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. Jaco replaces the function of the upper limb on one side of the body.

Lot/Code Info: Model Numbers: PJ 0000 0001, PJ 0000 0012, PJ 0090 0001, PJ 0090 0006, KR MJ2 0001. All lot numbers.

Quantity Affected: 949 units

Reason for Recall

Potential installation issue that could result in a fire hazard in the event of an interaction between a damaged Jaco arm and a wheelchair exhibiting electrical leakage.

Distribution

Worldwide - US Nationwide distribution in the states of California, Missouri, North Carolina, Ohio and the countries of Australia, Canada, China, Czechia, France, Germany, Netherlands, Norway, Portugal, Switzerland, Great Britain, Northern Ireland.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-08-22

Company

Kinova Inc

Montreal

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 180 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Kinova Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Kinova Inc have FDA actions?

This is the only FDA action we have on record for Kinova Inc in our database.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0049-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions