Class II Recall

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configurati

Canon Medical System, USA, INC.

Summary

The FDA issued a Class II for This device is a digital radiography/fluoroscopy system used in a diagnostic an by Canon Medical System, USA, INC.. Reason: It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and.

Details

Source

Device Recall

External ID

Z-0037-2026

Action Date

2025-10-15

Status

Ongoing

Category

device

Product Description

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Lot/Code Info: Serial Numbers: SZA2422001, SXA2432001, SUC23Z2014, BBF2392003, SUC2392013, SZA2422002, SXA2432002, SZA2442004.

Quantity Affected: 8

Reason for Recall

It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.

Distribution

U.S.

Type: FDA Mandated

Recall Initiated: 2025-09-02

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 161 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Canon Medical System, USA, INC. has 25 FDA actions in our database, including 25 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Canon Medical System, USA, INC.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Canon Medical System, USA, INC. have FDA actions?

Canon Medical System, USA, INC. has 25 FDA actions in our database, including 25 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0037-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.