RecallHawk
Class II Recall

Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical intervent

Getinge Usa Sales Inc

Summary

The FDA issued a Class II for Meera Mobile Operating Table-to support and position the patient immediately bef by Getinge Usa Sales Inc. Reason: Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates.

Details

Source

Device Recall

External ID

Z-0036-2024

Action Date

2023-10-11

Status

Ongoing

Category

device

Product Description

Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Model Number: 7200.01F2

Lot/Code Info: UDI Code: 04046768124363 Serial Numbers: 7 10 9 18 16 14 15 17 13 24 21 22 23 20 25 19 26 32 28 29 30 31 27 42 44 45 46 47 48 49 50 51 54 59 60 52 53 56 57 63 67 74 64 68 72 73 65 69 70 71 66 75 77 78 80 81 82 79 89 83 84 85 86 87 88 90 91 93 92 94 95 96 104 105 99 101 100 102 98 106 107 108 97 103 113 111 112 114 115 119 110 116 118 125 127 128 132 129 130 131 123 124 126 138 134 135 137 139 141 142 143 140 154 146 156 145 147 149 148 150 151 155 161 165 166 164 167 169 171 170 157 162 159 158 163 168 172 174 175 178 179 176 177 152 153 180 181 183 185 182 186 193 196 188 195 189 192 194 190 502 503 504 505 506 507 510 508 509 511 512 513 514 516 520 524 523 521 527 518 519 534 535 536 532 533 541 538 539 537 540 544 545 546 547 548 549 553 550 551 554 555 557 558 556 561 563 564 565 566 569 571 572 567 573 581 579 580 583 585 586 587 584 588 589

Quantity Affected: 224 units

Reason for Recall

Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. may result in the following reasonably foreseeable injury or damage to the health of patients and/or users

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2023-09-06

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 142 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Getinge Usa Sales Inc has 48 FDA actions in our database, including 48 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Getinge Usa Sales Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Getinge Usa Sales Inc have FDA actions?

Getinge Usa Sales Inc has 48 FDA actions in our database, including 48 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-0036-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions