Summary

NHTSA campaign 21V183: Braun Ambulances recalled for a safety defect. If the locking pin fails, the occupant may not be sufficiently restrained during a crash, increasing the risk of an inju

Details

Source

nhtsa

External ID

21V183

Action Date

2021-03-18

Status

Ongoing

Category

vehicle

Product Description

Braun Ambulances is recalling certain 2019 and 2021 MX170E Ford E-Series, and 2019-2020 MXP170E Dodge Ram 5500 Demers Ambulance vehicles equipped with a sliding attendant seat on the passenger-side of the ambulance patient compartment. The locking pin assembly in the sliding attendant seat may fail under certain loading conditions. As such, these vehicles fail to comply with the requirements of Federal Motor Vehicle Safety Standard (FMVSS) numbers 207, "Seating Systems" and 210, "Seat Belt Assembly Anchorages."

Context & Analysis

Braun Ambulances has 10 FDA actions in our database.

Frequently Asked Questions

How often does Braun Ambulances have FDA actions?

Braun Ambulances has 10 FDA actions in our database, including 0 recalls and 0 device clearances.